GXL-D500 ULTRA

 

 

CLINICAL OUTPUT

 

TESTIMONIAL

 

DOWNLOAD CATALOGUE

 

TECHNOLOGY

LF Generator Series offers advanced high quality Radiographic capabilities with unmatched performance and reliability resulting in the lowest cost of ownership over the product lifetime. This Generator is suitable for Film and CR - Based Radiographic systems and features leading edge micro-processor based controls to minimize patient doze and maximize image quality while demonstrating excellent reproducibility with user-friendly operator controls.

 

SALIENT FEATURES

  • Micro-processor based, Easy-Operational and User-Friendly X-Ray Control Console.
  • Micro-Processor controlled input line voltage compensation.
  • Modular and Smart Card based design for easy service.
  • Teflon wires used for Internal wiring for protection against Heat and Fire.
  • Powerful Circuit design provides maximum mA & kVp output accuracy.
  • Micro-Processor based real time exposure.
  • Exposure Trigger circuit control by SCR / TRIAC for soft firing.
  • Dual Exposure Switch. (Wired Hand Switch & Feather Touch Switches on Control Console)
  • Audible Tone at time of X-Ray Exposure.
  • Miniature Circuit Breaker protects against overload.
  • CVT with Electronic Stabilizer Card for mA circuit.
  • Micro-Processor protects X-Tube life from overload due to exposure factor selection automatically.
  • Precise selection of mAs brings better quality and reduces radiation time.
  • Auto shut down facilities (machine will automatically switch off after 30 min. if it is not in use).
  • Two options are available for Console : Stand Mounting & Wall Mounting.
 

TECHNICAL SPECIFICATION

 

MODEL NAME

GXL-D500

POWER SUPPLY REQUIREMENT

3 Phase

INPUT RATING

400 - 440 VAC / 50 KVA / 50Hz + 10%

LINE RESISTANCE

0.2 Ohms

OUTPUT RATING

40 kW

CONTROL PANEL

Microprocessor controlled feather touch operated

RAD kVp RANGE

35 – 125 kVp (Step of 1kVp and 10kVp)

RAD mAs RANGE

2 – 500 mAS (Step of 1mAs and 10mAs)

RAD mA RANGE

Small Focus - 100mA, 200mA

Large Focus - 300mA, 400mA, 500mA

RAD EXPOSURE TIME

0.01 – 5.0 Sec.

FLUORO kVp RANGE

35 – 100 kVp

FLUORO mA RANGE

0 – 3 mA Variable

FLUORO EXPOSURE TIME

0 – 300 Sec. Cumulative

DIGITAL DISPLAY

KV, mA, mAs, Timer (Displayed on Control Panel)

X-RAY TUBE

Rotation Anode Dual Focus

TUBE TYPE

Shell Type

RECTIFICATION

Full Wave Rectification

X-RAY FOCAL SPOT

1.0 MM2 Small, 2.0 MM2 Large

FLOOR TO CEILING TUBE STAND

Column stand for X-Ray Tube with movements covering full length of  X-Ray Table.

TABLE

Multi position Hand Tilt 5-Position Table

GRID

Motorized Bucky with moving Grid Ratio 8:1 size / 85Lines, enough to cover 17 1/4” X 18 7/8” Cassettes can move through whole length of table.

PHASE CONVERTER

3 to 1 transformer for Converting 3 Phase to 1 Phase power supply

H. V. CABLE

H.V. Cables 1 Pair

H. T. TRANSFORMER

H.T. Transformer for suitable Generator with stand

ACCESSORIES

Stainless Steel Cassette Tray, Compression Band, Foot Rest, Cone, Chest Stand, Multi Folding Screen, Hand Grip